Logo
User: Guest  Login
Authors:
Osinga, J.-M.; Akselrod, M. S.; Herrmann, Roland; Hable, Volker; Dollinger, Günther; Jäkel, O.; Greilich, S. 
Document type:
Zeitschriftenartikel / Journal Article 
Title:
High-accuracy fluence determination in ion beams using fluorescent nuclear track detectors 
Title of conference publication:
Proceedings of the 8th International Conference on Luminescent Detectors and Transformers of Ionizing Radiation (LUMDETR 2012) 
Journal:
Radiation Measurements 
Volume:
56 
Conference title:
International Conference on Luminescent Detectors and Transformers of Ionizing Radiation (8., 2012, Halle) 
Venue:
Halle 
Year of conference:
2012 
Year:
2013 
Pages from - to:
294-298 
Language:
Englisch 
Keywords:
Aluminum oxides ; Confocal laser scanning microscopy ; Heavy ion radiotherapy ; Laser scanning microscopy ; Linear energy transfer ; Nuclear track detector ; Radiation imaging ; Semi-automatic tools, Energy transfer ; Fluorescence, Ion beams 
Abstract:
We present an approach to use Al2O3:C; Mg-based fluorescent nuclear track detectors (FNTDs) and confocal laser scanning microscopy as a semiautomatic tool for fluence measurements in clinical ion beams. The method was found to cover a linear energy transfer (LET) range from at least L∞(Al2O3) = 0.5 keV/μm to 61; 000 keV/μm with a detection efficiency ≥99.83 
ISSN:
1350-4487 
Department:
Fakultät für Luft- und Raumfahrttechnik 
Institute:
LRT 2 - Institut für angewandte Physik und Messtechnik 
Chair:
Dollinger, Günther 
Open Access yes or no?:
Nein / No