Msimanga, M.; Comrie, C. M.; Pineda-Vargas, C. A.; Murray, S.; Bark, R.; Dollinger, Günther
Dokumenttyp:
Zeitschriftenartikel / Journal Article
Titel:
A Time of Flight-Energy spectrometer for stopping power measurements in Heavy Ion-ERD analysis at iThemba LABS
Zeitschrift:
Nuclear Instruments and Methods in Physics Research Section B
Jahrgang:
267
Heftnummer:
16
Jahr:
2009
Seiten von - bis:
2671-2674
Sprache:
Englisch
Stichwörter:
Detector systems ; Elastic recoil detection ; Energy loss ; Energy spectrometer ; Experimental data ; Quantitative analysis ; Stopping power ; Target matrices ; Thin layers ; Time of Flight, Electric power measurement ; Electron energy loss spectroscopy ; Energy dissipation ; Heavy ions ; Krypton ; Position control, Ionization
Abstract:
The quantitative analysis of thin layers using Heavy Ion-Elastic Recoil Detection (HI-ERD) can be reliably performed if the stopping powers of the probing ions and recoils in a given target matrix are known accurately. Unfortunately for many projectile/target combinations experimental data is limited and where available, deviations of up to 50